Philips HEARTSTART XLT M3500B Manuale Utente
System Messages
12-5
12
Momentary Messages
Momentary messages are temporary and only appear on the display for a few
seconds. Each message is accompanied by a three second beep to alert you.
Table 12-2 lists momentary messages.
seconds. Each message is accompanied by a three second beep to alert you.
Table 12-2 lists momentary messages.
Table 12-2 Momentary Messages
Message
Possible Cause
Corrective Action
$WWDFK3DGV
The multifunction defib electrode pads
are not making proper contact with the
patient.
are not making proper contact with the
patient.
z
Check the pads are applied to the
patient, as directed on the pads’ pack-
age.
patient, as directed on the pads’ pack-
age.
z
Replace the pads if the prompt contin-
ues.
ues.
$WWDFK/HDGV
An attempt was made to begin pacing in
Demand Mode without ECG leads
attached to the patient.
Demand Mode without ECG leads
attached to the patient.
Attach leads to patient.
'HILE'LVDUPHG
z
The pads connection is compromised.
z
The mode is changed from Manual to
AED while the defibrillator is
charged.
AED while the defibrillator is
charged.
z
is not pressed within 30 sec-
onds of the defibrillator being
charged.
charged.
z
is pressed.
z
Check the pads are applied to the
patient properly.
patient properly.
z
If a shock is indicated, deliver the
shock before changing modes.
shock before changing modes.
z
To deliver a shock, press
within 30 seconds of the defibrillator
being charged.
being charged.
1R6KRFN'HOLYHUHG
Patient impedance is too high.
z
Make sure the pads are applied prop-
erly.
erly.
z
Replace the pads, if necessary
&KHFN3ULQWHU
Printer paper is absent or jammed; the
printer door is not closed properly.
printer door is not closed properly.
z
Reload printer paper.
z
Make sure the door is closed properly.
6+2&.
',6$50
6+2&.