Philips HEARTSTART XLT M3500B Manuale Utente

Pagina di 180
System Messages
12-5
12
Momentary Messages
Momentary messages are temporary and only appear on the display for a few 
seconds. Each message is accompanied by a three second beep to alert you. 
Table 12-2 lists momentary messages.
Table 12-2  Momentary Messages
Message
Possible Cause
Corrective Action
$WWDFK3DGV
The multifunction defib electrode pads 
are not making proper contact with the 
patient.
z
Check the pads are applied to the 
patient, as directed on the pads’ pack-
age. 
z
Replace the pads if the prompt contin-
ues.
$WWDFK/HDGV
An attempt was made to begin pacing in 
Demand Mode without ECG leads 
attached to the patient. 
Attach leads to patient.
'HILE'LVDUPHG
z
The pads connection is compromised.
z
The mode is changed from Manual to 
AED while the defibrillator is 
charged.
z
 is not pressed within 30 sec-
onds of the defibrillator being 
charged.
z
 is pressed.
z
Check the pads are applied to the 
patient properly.
z
If a shock is indicated, deliver the 
shock before changing modes.
z
To deliver a shock, press 
 
within 30 seconds of the defibrillator 
being charged.
1R6KRFN'HOLYHUHG
Patient impedance is too high.
z
Make sure the pads are applied prop-
erly.
z
Replace the pads, if necessary
&KHFN3ULQWHU
Printer paper is absent or jammed; the 
printer door is not closed properly.
z
Reload printer paper. 
z
Make sure the door is closed properly.
6+2&.
',6$50
6+2&.