Performance Procedures
4-32 
Testing the Monitor  
Leakage
Check
The Leakage Check consists of two parts:
Part 1 - Checking the tubing between the pump outlet and the 
CO
module outlet.
Part 2 - Checking the tubing between the pump inlet and the FilterLine inlet.
Note
Check the flowmeter’s user guide for details on how to make a correct flow reading.
Part 1
Part 2
Step
Action
1
Check the ambient and cell pressure shown in the status line on the screen. The 
cell pressure should be approximately 20 mmHg (2.7 kPa) lower than the 
ambient pressure.
2
Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting 
tube.
3
Block  the  CO
module outlet using your fingertip and observe the flowmeter 
display. The value on the flowmeter (X2) should decrease to between 0 and 4 ml/
minute. If the value is not within the tolerance limits, there is a leakage between 
the pump outlet and the CO
module gas outlet. 
4
If a leakage is found in Step 3, open the CO
module and check the tubing 
connections at the pump outlet and the module CO
gas outlet. If the connections 
are good, then there is leakage in the module and the CO
module must be 
exchanged.
Step
Action
1
Disconnect the flowmeter from the Part 1 setup and connect the flowmeter inlet 
to the CO
module gas outlet.
2
Leave the FilterLine connected to the CO
module inlet.
3
Block the inlet of the FilterLine using your fingertip and observe the flowmeter 
display. The value on the flowmeter (X3) should decrease to between 0 and 4 ml/
minute. If the value is within the tolerance limits, there are no leakages and the 
leakage check is completed. You can proceed to the “Pump Check”.
4
If the value is not within the tolerance limits, there is a leakage between the 
FilterLine inlet and the pump inlet.
5
Check the FilterLine connections and open the monitor to check the tubing 
connections at the pump inlet and the CO
module gas inlet. If the connections 
are good, try replacing the FilterLine and repeating the leakage check. If the 
situation remains, there is a leakage in the tubing and the CO
module must be 
exchanged.