Philips 862474 ユーザーズマニュアル

ページ / 182
Performance Procedures
4-34 
Testing the Monitor  
 CO
2
 Gas
Measurement
Calibration
Check
After switching the monitor on, or after turning on the etCO
2
 On/Off setting in the etCO
2
 
frame, wait at least 20 minutes before checking the calibration. 
 
Step
Action
1
Check that the 5% calibration gas and flow regulator are connected.
2
Calculate the expected measurement value in mmHg as follows:
0.05 x (ambient pressure) / 1.03 = value mmHg
i.e. 0.05 x 736 mmHg / 1.03 = 35.7 mmHg (with an ambient pressure of 736 
mmHg)
Note: Dividing by 1.03 compensates for the dry calibration gas at room 
temperature, relative to breath gases containing water vapor and at body 
temperature.
3
Select CO
2
 Calibration. Wait for the status message "CO
2
 module reset in 
progress" to disappear. Open the valve on the flow regulator to allow 5% CO
2
 
gas to flow into the monitor. Allow the displayed CO
2
 value (third row of CO
2
 
Test menu) to stabilize.
4
Check that the CO
2
 value on the monitor matches the calculated mmHg value 
±2.6 mmHg. If the value is outside the tolerance, calibrate as described in Step 9 
to 13.
5
Disconnect the 5% calibration gas and connect the 10% calibration gas.
6
Calculate the expected measurement value and tolerance in mmHg as follows:
0.1 x (ambient pressure) / 1.03 = value mmHg
±0.07 x (value mmHg) = tolerance
i.e. 0.1 x 737 mmHg / 1.03 = 71.6 mmHg (with an ambient pressure of 737 
mmHg)
±0.07 x 71.6 mmHg = ±5.01 mmHg tolerance
7
Open the valve on the flow regulator to allow 10% CO
2
 gas to flow into the 
monitor. Allow the value to stabilize.
8
Check that the value on the monitor (X6) matches the calculated mmHg value 
within the calculated tolerance. If so, the CO
2
 module is correctly calibrated. If 
the value is outside the tolerance, calibrate as follows.
9
If not already connected, connect the 5% calibration gas.
10
Select CO
2
 Calibration to enter the Calibration sub-menu. Wait for the status 
message "CO
2
 module reset in progress" to disappear.
11
Select the value for the calibration gas (The default value is 5%).
12
Select Start Calibration and open the valve on the calibration gas to allow CO
2
 
gas to flow into the monitor. Allow the value to stabilize before the start of the 
calibration. Leave the valve open until the monitor gives a prompt that the gas 
can be removed.
13
The CO
2
 module calibrates and prompts when calibration is successful.