Philips 862474 ユーザーズマニュアル

ページ / 182
Performance Procedures
Testing the Monitor  
4-33
Pump Check
Flow Rate
Check and
Calibration
Note
If the adjusted flow is not stored within 60 seconds of the adjustment, the old flow setting is 
restored. If the flow cannot be adjusted to within tolerance, the pump should be replaced. If 
the flow adjustment still cannot be made, this indicates a fault in the 
CO
module, which must 
be replaced.
Step
Action
1
Connect the flowmeter inlet to the CO
module gas outlet.
2
Connect the FilterLine to the CO
module inlet.
3
Block the inlet of the FilterLine using your fingertip and observe the cell 
pressure to the right of the slash symbol on the top row of the CO
Test display 
screen.
The cell pressure (X4) should be more than 120 mmHg below the ambient 
pressure shown. If the pressure difference is less than 120 mmHg, the pump is 
not strong enough and should be replaced (regardless of the Pump Op Time).
Step
Action
1
Connect the flowmeter to the CO
FilterLine.
2
On the flowmeter, check the flow that the CO
Pump draws (X5). The nominal 
value is 50; the acceptable limits are 42.5 and 65 ml/minute. If the value is 
within these limits, proceed to the CO
2  
Gas calibration check. If not within the 
limits, calibrate as follows.
3
Select Flow Rate Check and Calibration. Adjust the flow in the monitor by 
selecting Increase Flow or Decrease Flow until it is as close as possible to 50 
ml per minute as indicated on the flowmeter gauge. The pump voltage displayed 
on the second row of the CO
Test screen will vary as the flow is changed.
4
When you are satisfied that the flow is set as close as possible to 50 ml per 
minute, select Store Flow to confirm the setting