Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
2-4
 
Making Mixer Measurements (Option 089 Only)
Measurement Considerations
Measurement Considerations 
In mixer transmission measurements, you have RF and LO inputs and an IF output. Also 
emanating from the IF port are several other mixing products of the RF and LO signals. In 
mixer measurements, leakage signals from one mixer port propagate and appear at the 
other two mixer ports. These unwanted mixing products or leakage signals can cause 
distortion by mixing with a harmonic of the analyzer’s first down-conversion stage.
To ensure successful mixer measurements, the following measurement challenges must be 
taken into consideration:
• Mixer Considerations
• Analyzer Operation
Minimizing Source and Load Mismatches 
When characterizing linear devices, you can use vector accuracy enhancement to 
mathematically remove all systematic errors, including source and load mismatches, from 
your measurement. This is difficult when the device you are characterizing is a mixer 
where the input and output signals are at different frequencies. Therefore, source and load 
mismatches are not corrected for and will add to overall measurement uncertainty.
You should place attenuators at all of the test ports to reduce the measurement errors 
associated with the interaction between mixer port matches and system port matches. To 
avoid overdriving the receiver, you should give extra care to selecting the attenuator 
located at the mixer's IF port to avoid overdriving the receiver. For best results, you should 
choose the attenuator value so that the power incident on the analyzer R channel input is 
less than 
10 dBm and greater than 
35 dBm.