Making Mixer Measurements (Option 089 Only)
Amplitude and Phase Tracking
Amplitude and Phase Tracking 
The match between mixers is defined as the absolute difference in amplitude or phase 
response over a specified frequency range. The tracking between mixers is essentially how 
well the devices are matched over a specified interval. This interval may be a frequency 
interval or a temperature interval, or a combination of both.
You can make tracking measurements by ratioing the responses of two mixer conversion 
loss measurements. Then any difference you view in response is due to the mixers and not 
the measurement system.
Using the same measurement setup as in 
you can determine how well two mixers track each other in terms of amplitude and 
phase.
1. Repeat steps 1 through 8 of the previous section 
, with the following exception:
In step 7, select 
 
.
2. Once the analyzer has displayed the measurement results, press 
 
.
3. Replace the calibration mixer with the mixer under test.
4. Press .
The resulting trace should represent the amplitude and phase tracking of the two mixers.
Figure 2-26 Connections for an Amplitude and Phase Tracking Measurement 
Between Two Mixers
Format
 PHASE 
Display
 DATA
MEM 
 DATA/MEM