Making Mixer Measurements (Option 089 Only)
Conversion Compression Using the Frequency Offset Mode
Conversion Compression Using the Frequency Offset Mode 
Conversion compression is a measure of the maximum RF input signal level where the 
mixer provides linear operation. The conversion loss is the ratio of the IF output level to 
the RF input level. This value remains constant over a specified input power range. When 
the input power level exceeds a certain maximum, the constant ratio between IF and RF 
power levels will begin to change. The point at which the ratio has decreased by 1 dB is 
called the 1 dB compression point. See 
Figure 2-27 Conversion Loss and Output Power as a Function of Input Power 
Level Example 
Notice that the IF output power increases linearly with the increasing RF signal, until 
mixer compression begins and the mixer saturates.
The following example uses a ratio of mixer output to input power and a marker search 
function to locate a mixer’s 1 dB compression point.
NOTE
Because this procedure was performed with an 8719ES/20ES, Option 007, the 
analyzer was able to produce an output power of +10 dBm.
If the default output power of your analyzer is not high enough to force the 
mixer under test into compression, then the following procedure may have to 
be performed with the addition of Option 007 or Option 085. Refer to the 
.
1. Set the LO source to the desired CW frequency of 600 MHz and power level to 13 dBm.
2. Initialize the analyzer by pressing 
.
3.  Set the analyzers LO frequency to match the frequency of the LO source by pressing:
 
 
Preset
Sweep Setup
 SWEEP TYPE MENU   POWER SWEEP 
Return