Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
3-26
 
Making Time Domain Measurements
Masking
Masking 
Masking occurs when a discontinuity (fault) closest to the reference plane affects the 
response of each subsequent discontinuity. This happens because the energy reflected from 
the first discontinuity never reaches subsequent discontinuities. For example, if a 
transmission line has two discontinuities that each reflect 50% of the incident voltage, the 
time domain response (real format) shows the correct reflection coefficient for the first 
discontinuity (
ρ
=0.50). However, the second discontinuity appears as a 37.5% reflection 
(
ρ
=0.375) because only some the incident voltage reached the second discontinuity, and 
some of that reflected energy is reflected off the first discontinuity as it returns. For two 
discrete discontinuities, the apparent reflection of the second discontinuity is appears as 
approximately 
, where 
 is the apparent reflection of the second 
discontinuity, 
 is the reflection of the first discontinuity, and 
 is the reflection of the 
second discontinuity.
NOTE
This example assumes a lossless transmission line. Real transmission lines, 
with non-zero loss, attenuate signals as a function of the distance from the 
reference plane.
As an example of masking due to line loss, consider the time domain response of a 3 dB 
attenuator and a short circuit. The impulse response (log magnitude format) of the short 
circuit alone is a return loss of 0 dB, as shown in 
a. When the short circuit is 
placed at the end of the 3 dB attenuator, the return loss is 
6 dB, as shown in 
b. 
This value actually represents the forward and return path loss through the attenuator, 
and illustrates how a lossy network can affect the responses that follow it.
Figure 3-21 Masking Example 
ρ
ˆ
2
1
ρ
1
2
(
) ρ
2
=
ρ
ˆ
2
ρ
1
ρ
2