Calibrating for Increased Measurement Accuracy
Calibration Considerations
Figure 6-1 Typical Responses of Calibration Standards after Calibration 
Interpolated Error Correction 
You may want to use interpolated error correction when you choose a subset of a frequency 
range that you already corrected, when you change the number of points, or when you 
change to CW. This feature also allows you to change the parameters in a 2-port correction, 
such as IF bandwidth, power, or sweep time. The analyzer calculates the systematic errors 
from the errors of the original correction.
To activate interpolated measurement correction, press 
 
 so that 
ON is selected, and 
 so that ON is selected. When interpolation is 
in use, the notation 
C
 will appear on the analyzer display.
1
1
1
1
START
START
START
START
CH1 S11
1 U FS
CH1 S11
1 U FS
CH1 S11
1 U FS
CH1 S11
1 U FS
1: 998.84 mU
-179.92
°
1: 998.25 mU
-11.779
°
1: 999.62 mU
142.07
°
1: 999.1 mU
-44.561
°
STOP 3
STOP 3
STOP 3
STOP 3
.300 000 MHZ
.300 000 MHZ
.300 000 MHZ
.300 000 MHZ
000 000 000 MHZ
.
000 000 000 MHZ
.
000 000 000 MHZ
.
000 000 000 MHZ
.
3 000 000 000 MHz
.
3 000 000 000 MHz
.
3 000 000 000 MHz
.
3 000 000 000 MHz
.
Cor
Cor
Cor
Cor
7mm or Type-N Male
Short (No Offset)
7mm or Type-N Male
Open (No Offset) with Fringing Capacitance
Type-N Female,
3.5mm Male or Female Offset Short
Type-N Female
3.5 mm Male or Female Offset Open
pa5162e
Cal
 INTERPOL ON off 
CORRECTION on OFF