Calibrating for Increased Measurement Accuracy
Calibration Considerations
NOTE
The preset state of the instrument can be configured so that interpolated 
error correction is on or off. Press 
 
 
 
 
 to configure 
the preset state of interpolated error correction.
System performance is unspecified when using interpolated error correction. The quality of 
the interpolated error correction is dependent on the amount of phase shift and the 
amplitude change between measurement points. If phase shift is no greater than 180
°
 per 
approximately five measurement points, interpolated error correction offers a great 
improvement over uncorrected measurements. The accuracy of interpolated error 
correction improves as the phase shift and amplitude change between adjacent points 
decrease. When you use the analyzer in linear frequency sweep, perform the original 
calibration with at least 30 points per 1 GHz of frequency span for greatest accuracy with 
interpolated error correction.
Interpolated error correction is available in three sweep modes: linear frequency, power 
sweep, and CW time.
NOTE
If there is a valid correction array for a linear frequency sweep, this may be 
interpolated to provide correction at the CW frequency used in power sweep 
or CW time modes. This correction is part of the interpolated error-correction 
feature.
Error-Correction Stimulus State 
Error correction is only valid for a specific stimulus state, which you must select before you 
start a correction. If you change any of the following parameters, you will invalidate the 
correction and the analyzer will switch the correction off (unless the interpolated error 
correction feature is activated):
• frequency range
• number of points
• sweep type
The error correction quality may be degraded (
Cor
 changes to 
C
), if you change the 
following stimulus state parameters:
• sweep time
• system bandwidth
• output power
If correction is turned off or shows 
C
, the original stimulus state can be recalled by, first, 
turning interpolation off (
), and toggling correction off and then on 
(
).
System
 CONFIGURE MENU 
 USER SETTINGS   PRESET SETTINGS   CAL INTERP ON off 
 INTERPOL ON off 
 CORRECTION ON off