Calibrating for Increased Measurement Accuracy
Procedures for Error Correcting Your Measurements
Procedures for Error Correcting Your Measurements 
This section has example procedures or information on the following topics:
• frequency response correction
• frequency response and isolation correction
• enhanced frequency response correction (with enhanced reflection error correction)
• one-port reflection correction
• full two-port correction (ES Analyzers Only)
• TRL*/LRM* correction (ES Analyzers Only)
• power meter measurement calibration procedure
NOTE
If the channels are uncoupled, you must make a correction for each channel.
Types of Error Correction 
Several types of error correction are available that remove from one to twelve systematic 
errors. The full 2-port correction effectively removes all twelve correctable systematic 
errors. Some measurements do not require correction for all twelve errors
explains each correction and its uses.
Table 6-2  Purpose and Use of Different Error Correction Procedures 
Correction Procedure 
Corresponding 
Measurement 
Errors Corrected 
Standard Devices 
Response
Transmission or 
reflection 
measurement when 
the highest accuracy 
is not required.
Frequency response.
Thru for 
transmission, open or 
short for reflection.
Response & Isolation
Transmission of high 
insertion loss devices 
or reflection of high 
return loss devices. 
Not as accurate as 
1-port or 2-port 
correction.
Frequency response 
plus isolation in 
transmission or 
directivity in 
reflection.
Same as response 
plus isolation 
standard. (load)