6-12
 
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
Frequency Response Error Corrections 
You can remove the frequency response of the test setup for the following measurements:
• reflection measurements
• transmission measurements
• combined reflection and transmission measurements
Response Error Correction for Reflection Measurements 
1. Press .
2. Select the type of measurement you want to make.
❏ If you want to make a reflection measurement on PORT 1 (in the forward direction, 
S
11
), leave the instrument default setting.
❏ For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the 
reverse direction, S
22
), press:
 
3. Set any other measurement parameters that you want for the device measurement: 
power, sweep type, number of points, or IF bandwidth.
4. To access the measurement error-correction menus, press:
 
5. If your calibration kit is different than the kit specified under the 
 
 
softkey, press:
 
 
 (select your type of kit) 
If your type of calibration kit is not listed in the displayed menu, refer to 
6. To select a response correction, press:
 
 
Connect the short or open calibration standard to the port you selected for the test port 
(PORT 1 for S
11
 or PORT 2 for S
22
 on ES analyzers).
NOTE
Include any adapters or cables that you will have in the device measurement. 
That is, connect the standard device to the particular connector where you 
will connect your DUT.
Preset
Meas
 Refl: REV S22 (B/R) 
Cal
 CALIBRATE MENU 
Cal
 CAL KIT [ ] 
Cal
 CAL KIT   SELECT CAL KIT 
 RETURN 
Cal
 CALIBRATE MENU   RESPONSE