6-14
 
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
Response Error Correction for Transmission Measurements 
1. Press .
2. Select the type of measurement you want to make.
❏ If you want to make a transmission measurement in the forward direction (S
21
), 
press:
 
 or on ET models: 
❏ For ES analyzers, if you want to make a transmission measurement in the reverse 
direction (S
12
), press:
 
3. Set any other measurement parameters that you want for the device measurement: 
power, number of points, IF bandwidth.
4. To select a response correction, press:
 
 
5. Make a "thru" connection between the points where you will connect your DUT.
NOTE
Include any adapters or cables that you will have in the device measurement. 
That is, connect the standard device where you will connect your DUT.
Figure 6-3 Standard Connections for Response Error Correction for 
Transmission Measurements
6. To measure the standard, press:
The analyzer displays 
WAIT - MEASURING CAL STANDARD
 during the standard 
measurement. The analyzer underlines the 
 softkey after it measures the 
calibration standard, and computes the error coefficients.
Preset
Meas
 Trans: FWD S21 (B/R) 
 TRANSMISSN 
Meas
 Trans: REV S12 (A/R) 
Cal
 CALIBRATE MENU   RESPONSE 
 THRU 