Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
6- 15
Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
NOTE
Do not use an open or short standard for a transmission response correction.
NOTE
You can save or store the measurement correction to use for later 
measurements. Refer to the 
 for procedures.
7. This completes the response correction for transmission measurements. You can 
connect and measure your device under test.
Receiver Calibration 
Receiver calibration provides a frequency response error correction for a non-ratioed 
measurement that also indicates absolute power in dBm. This calibration is most useful 
when performed with a power meter calibration. This calibration is only allowed for 
non-ratioed measurements A, B, and R.
This calibration normalizes the trace to the current reference value. Typically, this 
reference value is entered to be the same as the current source power.
1. Perform a power meter calibration to the desired level. Refer to step A of 
Use 
10 dBm for this example. (See also, 
.) This provides a calibrated power, referenced to the power meter, to use as a 
receiver calibration standard.
or
Set the analyzer test port power to the desired level (
10 dBm in this example) by 
pressing:
 
 
. This calibrates the receiver to the approximate accuracy of the 
source output power, which is subject to the source power flatness specification.
2. Make a "thru" connection between the points where you will connect your device under 
test. Refer to Step B of 
.
NOTE
Include any adapters or cables that you will have in the device measurement. 
That is, connect the standard device where you will connect your device under 
test.
Power
10
x1