Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
Using Continuous Correction Mode 
You can set the analyzer to update the correction table at each sweep (as in a leveling 
application), using the continuous sample mode. When the analyzer is in this mode, it 
continuously checks power at every point in each sweep. You must keep the power meter 
connected as shown in 
. This mode is also known as power meter leveling, and 
the speed is limited by the power meter.
NOTE
You may level at the input of a device under test, using a 2-resistor power 
splitter or a directional coupler before the device; or level at the output of the 
device using a 3-resistor power splitter or a bidirectional coupler after the 
device.
Figure 6-11 Continuous Correction Mode for Power Meter Calibration
1. Connect a power splitter or directional coupler to the port supplying RF power to your 
test device, as shown in 
.
2. Set test port power to approximate desired leveled power.
3. Press 
 
 and enter the test port power level that you want the 
analyzer to maintain at the input to your test device. Compensate for the power loss of 
the power splitter or directional coupler in the setup.
4. If you want the analyzer to make more than one power measurement at each frequency 
data point, press 
 
 
 (where n = the number of desired 
iterations).
If you increase the number of readings, the power meter correction time will 
substantially increase.
5. Press 
 
 
 
 to activate the 
power meter correction.
Cal
 PWRMTR CAL 
 NUMBER OF READINGS 
n
x1
Cal
 PWRMTR CAL   EACH SWEEP   TAKE CAL SWEEP