Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
7-14
 
Operating Concepts
Channel Stimulus Coupling
Channel Stimulus Coupling 
 toggles the channel coupling of stimulus values. With 
 (the preset condition), both channels have the same stimulus values. 
(The inactive channel takes on the stimulus values of the active channel.)
In the stimulus coupled mode, the following parameters are coupled:
• frequency
• number of points
• source power
• number of groups
• IF bandwidth
• sweep time
• trigger type
• gating parameters
• sweep type
• power meter calibration
Coupling of stimulus values for the two channels is independent of 
 
in the display menu and 
 in the marker mode menu. 
 activates an alternate sweep function when dual channel display is 
on. In this mode, the analyzer alternates between the two sets of stimulus values and 
displays the measurement data of both channels.
COUPLED CH on OFF 
COUPLED CH ON 
DUAL CHAN on OFF 
MARKERS: UNCOUPLED 
COUPLED CH OFF