Operating Concepts
Sweep Types
Sweep Types 
The following sweep types will function with the interpolated error-correction feature 
(described in 
• linear frequency
• power sweep
• CW  time
The following sweep types will not function with the interpolated error correction feature:
• logarithmic frequency sweep
• list frequency sweep
Linear Frequency Sweep (Hz) 
The 
 softkey activates a linear frequency sweep that is displayed on a standard 
graticule with ten equal horizontal divisions. This is the preset default sweep type.
For a linear sweep, sweep time is combined with the channel's frequency span to compute a 
source sweep rate:
         sweep rate = (frequency span) / (sweep time)
Since the sweep time may be affected by various factors, the equation provided here is 
merely an indication of the ideal (fastest) sweep rate. If the user-specified sweep time is 
greater than 15 ms times the number of points, the sweep changes from a continuous ramp 
sweep to a stepped CW sweep. Also, for 10 Hz or 30 Hz IF bandwidths, the sweep is 
automatically converted to a stepped CW sweep.
In the linear frequency sweep mode, it is possible, with Option 010, to transform the data 
for time domain measurements using the inverse Fourier transform technique.
Logarithmic Frequency Sweep (Hz) 
The 
 softkey activates a logarithmic frequency sweep mode. The source is 
stepped in logarithmic increments and the data is displayed on a logarithmic graticule. 
This is slower than a continuous sweep with the same number of points, and the entered 
sweep time may therefore be changed automatically. For frequency spans of less than two 
octaves, the sweep type automatically reverts to linear sweep.
Stepped List Frequency Sweep (Hz) 
The 
 softkey activates a stepped list frequency sweep, one of two 
list frequency sweep modes. The stepped list mode allows the analyzer to sweep a list of 
arbitrary frequency points. This list is defined and modified using the edit list menu and 
the edit subsweep menu. Up to 30 frequency subsweeps (called "segments") of several 
different types can be specified, for a maximum total of 1601 points. 
LIN FREQ 
LOG FREQ 
LIST FREQ [STEPPED]