Operating Concepts
Analyzer Display Formats
Smith Chart Format 
The 
 softkey displays a Smith chart format. Refer to 
. This is 
used in reflection measurements to provide a readout of the data in terms of impedance. 
The intersecting dotted lines on the Smith chart represent constant resistance and 
constant reactance values, normalized to the characteristic impedance, Z
0
, of the system. 
Reactance values in the upper half of the Smith chart circle are positive (inductive) 
reactance, and those in the lower half of the circle are negative (capacitive) reactance. The 
default marker readout is in ohms (
Ω
) to measure resistance and reactance (R+jX). 
Additional marker types are available in the Smith marker menu.
The Smith chart is most easily understood with a full scale value of 1.0. If the scale per 
division is less than 0.2, the format switches automatically to polar.
If the characteristic impedance of the system is not 50 ohms, modify the impedance value 
recognized by the analyzer by pressing 
 
 
 (the impedance value) 
.
An inverted Smith chart format for admittance measurements is also available. This is 
shown in 
. Access this by selecting 
 in the format menu, and 
pressing  
 
 . 
The 
Smith chart is inverted and marker values are read out in siemens (S) to measure 
conductance and susceptance (G+jB).
Figure 7-9 Standard and Inverse Smith Chart Formats 
SMITH CHART 
Cal
MORE  SET Z0 
x1
SMITH CHART 
Marker Fctn
MKR MODE MENU  SMITH MKR MENU  G+jB MKR