Operating Concepts
Analyzer Display Formats
Polar Format 
The 
 softkey displays a polar format as shown in 
. Each point on the 
polar format corresponds to a particular value of both magnitude and phase. Quantities 
are read vectorally: the magnitude at any point is determined by its displacement from the 
center (which has zero value), and the phase by the angle counterclockwise from the 
positive x-axis. Magnitude is scaled in a linear fashion, with the value of the outer circle 
usually set to a ratio value of 1. Since there is no frequency axis, frequency information is 
read from the markers.
The default marker readout for the polar format is in linear magnitude and phase. A log 
magnitude marker and a real/imaginary marker are available in the polar marker menu.
Figure 7-10 Polar Format 
Linear Magnitude Format 
The 
 softkey displays the linear magnitude format as shown in 
This is a Cartesian format used for unitless measurements such as reflection coefficient 
magnitude 
ρ
 or transmission coefficient magnitude 
τ,
 and for linear measurement units. It 
is used for display of conversion parameters and time domain transform data.