Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
7-78
 
Operating Concepts
GPIB Operation
GPIB Operation 
This section contains information on the following topics:
• local key
• GPIB controller modes
• instrument addresses
• using the parallel port
 Key 
This key is allows you to return the analyzer to local (front panel) operation from remote 
(computer controlled) operation. This key will also abort a test sequence or hardcopy 
print/plot. In this local mode, with a controller still connected on GPIB, you can operate the 
analyzer manually (locally) from the front panel. This is the only front panel key that is 
not disabled when the analyzer is remotely controlled over GPIB by a computer. The 
exception to this is when local lockout is in effect: this is a remote command that disables 
the 
 key, making it difficult to interfere with the analyzer while it is under 
computer control.
In addition, the 
 key accesses the GPIB menu, where you can set the controller 
mode, and to the address menu, where you can enter the GPIB addresses of peripheral 
devices and select plotter/printer ports. You can also set the mode of the parallel port here.
The GPIB menu consists of the following softkeys:
The analyzer is factory-equipped with a remote programming interface using the General 
Purpose Interface Bus (GPIB). This enables communication between the analyzer and a 
controlling computer as well as other peripheral devices. This menu indicates the present 
GPIB controller mode of the analyzer. Three GPIB modes are possible: system controller, 
talker/listener, and pass control.
Local
Local
Local
SYSTEM CONTROLLER 
TALKER/LISTENER 
USE PASS CONTROL 
SET ADDRESS 
PARALLEL [ ] 
GPIB DIAG on OFF 
DISK UNIT NUMBER 
VOLUME NUMBER