Operating Concepts
GPIB Operation
Address Menu 
This menu can be accessed by pressing the 
 softkey within the GPIB 
menu.
In communications through the General Purpose Interface Bus (GPIB), each instrument 
on the bus is identified by a GPIB address. This decimal-based address code must be 
different for each instrument on the bus.
This menu lets you set the GPIB address of the analyzer, and enter the addresses of 
peripheral devices so that the analyzer can communicate with them.
Most of the GPIB addresses are set at the factory and need not be modified for normal 
system operation. The standard factory-set addresses for instruments that may be part of 
the system are as follows:
The address displayed in this menu for each peripheral device must match the address set 
on the device itself. The analyzer does not have a GPIB switch: its address is set only from 
the front panel.
These addresses are stored in non-volatile memory and are not affected by preset or by 
cycling the power.
Using the Parallel Port 
The instrument’s parallel port can be used in two different modes. By pressing 
 and 
then toggling the 
 softkey, you can select either the 
 mode or the 
 mode.
The Copy Mode 
The copy mode allows the parallel port to be connected to a printer or plotter for the 
outputting of test results. To use the parallel port for printing or plotting, you must do the 
following:
1. Press 
 .
2. Select either 
 or 
.
3. Select 
 so that copy is underlined.
Instrument GPIB 
Address
(decimal) 
Analyzer
16
Plotter
05
Printer
01
External Disk Drive
00
Controller
21
Power Meter
13
SET ADDRESS 
Local
PARALLEL [   ] 
[COPY] 
[GPIO] 
Local
SET ADDRESSES 
PLOTTER PORT 
PRINTER PORT 
PARALLEL