Operating Concepts
GPIB Operation
GPIB STATUS Indicators 
When the analyzer is connected to other instruments over GPIB, the GPIB STATUS 
indicators in the instrument state function block light up to display the current status of 
the analyzer.
R = remote operation
L = listen mode
T = talk mode
S = service request (SRQ) asserted by the analyzer
System Controller Mode 
The 
 softkey activates the system controller mode. When in this 
mode, the analyzer can use GPIB to control compatible peripherals, without the use of an 
external computer. It can output measurement results directly to a compatible printer or 
plotter, store instrument states using a compatible disk drive, or control a power meter for 
performing service routines. The power meter calibration function requires system 
controller or pass control mode.
Talker/Listener Mode 
The 
 softkey activates the talker/listener mode, which is the mode of 
operation most often used. In this mode, a computer controller communicates with the 
analyzer and other compatible peripherals over the bus. The computer sends commands or 
instructions to and receives data from the analyzer. All of the capabilities available from 
the analyzer front panel can be used in this remote operation mode, except for control of 
the power line switch and some internal tests.
Pass Control Mode 
The 
 softkey activates the third mode of GPIB operation: the pass 
control mode. In an automated system with a computer controller, the controller can pass 
control of the bus to the analyzer on request from the analyzer. The analyzer is then the 
controller of the peripherals, and can direct them to plot, print, or store without going 
through the computer. When the peripheral operation is complete, control is passed back to 
the computer. Only one controller can be active at a time. The computer remains the 
system controller, and can regain control at any time.
Preset does not affect the selected controller mode, but cycling the power returns the 
analyzer to talker/listener mode.
Information on compatible peripherals is provided in the “Options and Accessories” 
chapter of the reference guide.
SYSTEM CONTROLLER 
TALKER/LISTENER 
USE PASS CONTROL