Graco 3A0869J ユーザーズマニュアル

ページ / 80
Basic Operation 
32
3A0869J
Prime the System
NOTE: SeRun Mode Details, pages 42-43, for further 
screen information, if needed.
1. Adjust the main air pressure. Most applications 
require about 80 psi (552 kPa, 5.5 bar) air pressure 
to operate properly. Do not use less than 75 psi 
(517 kPa, 5.2 bar).
2. If this is the first time starting up the system, or if 
lines may contain air, purge as instructed in Purg-
ing
, page 34. The equipment was tested with light-
weight oil, which should be flushed out to avoid 
contaminating your material.
3. From Run Home (Screen 1), press 
. Make 
sure that the system is in Standby mode.
4. Adjust component A and B fluid supplies 
as needed for your application. Use low-
est pressure possible.
NOTE: Do not exceed the maximum rated 
working pressure shown on the system identification 
label or the lowest rated component in the system.
5. Open the fluid supply valves to the 
system.
6. If using an electrostatic gun, shut 
off the electrostatics before spray-
ing.
7. If using a gun flush box, place the gun in the box 
and close the lid. 
8. Use 
 to change to desired color recipe.
9. Press 
. The system will purge, then load mixed 
material to the gun. If the gun flush box is not used, 
trigger the gun into a grounded metal pail until the 
system returns to Standby.
Meter Calibration
NOTE: See Calibration 1 and 2 (Screens 22 and 23)
page 49, for further screen information, if needed.
Calibrate the meter:
The first time the system is operated.
Whenever new materials are used in the system, 
especially if the materials have viscosities that differ 
significantly.
As part of regular maintenance to retain meter accu-
racy.
Whenever a flow meter is serviced or replaced.
NOTE: 
Meter factors on Calibration 1 (Screen 22) are 
updated automatically after the calibration pro-
cedure is completed. You also may manually 
edit them if desired.
All values on this screen are in cc or cc/pulse, 
regardless of the units set in Configure 4 
(Screen 21).
1. Before calibrating Meter A or Meter B, prime the 
system with material. See Prime the System, page 
32.
2. If the display is on a Run Mode screen, press 
 
to access setup screens.
3. Press 
 to display Calibration 1 (Screen 22). 
K-factors are shown for B and A1 (A2, A3 if pres-
ent).
Standby
Standby