CANON  Digital Galvano Scanner System  GM-1000 Series 
Users Manual 1.20 
 
 
 
  
68
 
Note: 
The analog monitor output is from the D/A converter of the controller. 
Since the output range of the D/A converter is from -3 to +3 V, output is repeated by the 
magnification setting if it exceeds this range. 
 
The relationship between encoder pulse and angle differs depending on the galvano motor. 
 E.g. 
GM-1010 
                360° = 1000 pulses × 8,192 divisions = 8,192,000 pulses 
        1° = 8,192,000 pulses / 360° = 22,756 pulses 
                      At the monitor output position (0.091 V/pulse), the analog monitor output changes: 
        1° = 22,756 × 0.091 = 2.071 V 
When the magnification is 1x, and there is a position change of 0° → 1°, analog output will 
change from 0V  → 2.071V. 
 
                E.g. GM-1015 , GM-1020 
                360° = 1500 pulses × 8,192 divisions = 12,288,000 pulses 
                1° = 12,288,000 pulses / 360 = 34,133 pulses 
                      At the monitor output position (0.091 V/pulse), the analog monitor output changes: 
        1° = 34,133×0.091 = 777 mV 
When the magnification is 1x, and there is a position change of 0° → 1°, analog output will 
change from 0V  → 777 mV. 
 
 
RS-232C Commands 
・ 
Monitor Output Selection (A2, A8)         
Command ID = 40   
・ 
Monitor Output Selection (A3, A9)        
Command ID = 41   
・ 
Monitor Output Selection (A4, A10)       
Command ID = 42   
・ 
Monitor Magnification Setting (A2, A8)       Command ID = 44   
・ 
Monitor Magnification Setting (A3, A9)     
Command ID = 45   
・ 
Monitor Magnification Setting (A4, A10)     
Command ID = 46