5 4 0 0 3 2 1 R e v . X : D R A F T
the stripper away from the handpiece in a smooth motion to ensure that the jacket is
cleanly removed.
cleanly removed.
5) Use a diamond/carbide cleaver to cut the used end of the fiber. Place the fiber
against a flat surface. Position the edge of the cleaver approximately ¼ inch from the
end of the fiber, and make a scratch around half the circumference of the fiber. Make
sure that the edge of the cleaver is always perpendicular to the fiber during
scratching.
end of the fiber, and make a scratch around half the circumference of the fiber. Make
sure that the edge of the cleaver is always perpendicular to the fiber during
scratching.
6) Hold the end of the fiber above the scratch between thumb and forefinger and pull the
end of the fiber away until the end section breaks off. If the fiber end is removed
properly by pulling in the direction perpendicular to the end surface of the fiber, the
fiber should end in a flat surface.
properly by pulling in the direction perpendicular to the end surface of the fiber, the
fiber should end in a flat surface.
7) Verify the cleave quality by aiming the fiber at a flat surface and observing the shape
of the spot created by the visible aiming beam. If the visible spot is a full circle, then
the power output is optimal; if the circle is distorted, then repeat only the cleaving
procedure presented in steps 5 and 6 until you obtain a perfect circle beam.
the power output is optimal; if the circle is distorted, then repeat only the cleaving
procedure presented in steps 5 and 6 until you obtain a perfect circle beam.
8) After the fiber is successfully cleaved, pull the fiber back through the handpiece until
just the fiber tip protrudes from the fiber end of the handpiece head. Tighten the
handpiece base until snug. Ensure that the fiber isn’t loose by pulling tightly on the
fiber optic cable at the proximal end.