Laser Parameters:
 
 
 
Tip Diameter (µm) 
Power (W) 
Mode 
400 
1.4 
CW 
300 
1.0 
CW 
 
■ 
Procedural Steps
 
Step 1)  
 
Set the EPIC 10 to the appropriate setting for the particular    
tip, using the table above as a guide.                                     
 
 
Step 2) 
 
Touch the tip to the surface of the initiation block, without firing
 
 
 
 
Step 3) 
 
Fire the laser, allowing the tip to sink into the block.  Pull the tip 
out when the metal canula touches the block, still firing until just 
before
 
 
 
 
Step 4) 
 
Fire the laser into the air once, you will see a white flash or 
the tip will glow.
 
 
 
 
 
Repeat steps 1 – 4 to ensure the tip is initiated. 