Biolase Inc. EPIC ユーザーズマニュアル
21 |
P a g e
5 4 0 0 3 2 1 R e v . X : D R A F T
■
Laser Parameters:
Tip Diameter (µm)
Power (W)
Mode
400
1.4
CW
300
1.0
CW
■
Procedural Steps
:
Step 1)
Set the EPIC 10 to the appropriate setting for the particular
tip, using the table above as a guide.
tip, using the table above as a guide.
Step 2)
Touch the tip to the surface of the initiation block, without firing
Step 3)
Fire the laser, allowing the tip to sink into the block. Pull the tip
out when the metal canula touches the block, still firing until just
before
out when the metal canula touches the block, still firing until just
before
Step 4)
Fire the laser into the air once, you will see a white flash or
the tip will glow.
the tip will glow.
Repeat steps 1 – 4 to ensure the tip is initiated.
Section 4: OPERATING INSTRUCTIONS