Biolase Inc. EPIC ユーザーズマニュアル

ページ / 45
23 | 
P a g e
                                                                             
5 4 0 0 3 2 1 R e v .   X :   D R A F T
 
 
       
            
 
                           Figure 21                                                   Figure 22
 
  Enter the three digit key access code using the touch screen.  The Access Key 
Code is 888. (If the proper code is not entered, re-enter the correct code.) 
  The system will go to the HOME screen, which identifies three procedure 
categories to choose from. 
  Verify wireless communication with the footswitch; a blue LED indicator light on 
the console will blink when communication is established.  The laser and 
footswitch are shipped already paired.  However, if the wireless communication is 
not confirmed by the indicator light, take the following steps to establish pairing: 
 
 
1.   Go to the Settings menu on the console display and select “Wireless” icon. 
2.   In the “Wireless” menu, press “Connect” to initiate pairing; blue LED will start  
blinking, indicating that the laser console and footswitch are communicating. 
3.  While holding the footswitch, press and hold the Reset switch on the bottom 
of the footswitch. 
4.  While holding the footswitch, press and hold the foot pedal. 
5.  Release the Rest switch. 
6.  Release the foot pedal. 
7.  A message will appear on the console display indicating that pairing has 
been successful, or unsuccessful.  If unsuccessful, repeat steps 1-6.  
 
CONTROL Button 
Pressing and holding the Control Button for 2 seconds will cause the system to enter 
SLEEP mode, or move out of SLEEP mode to STANDBY mode, or go from 
STANDBY mode to READY mode; this button will also turn the system ON. 
 
Entering READY or STANDBY Modes 
A quick press of the Control Button will place the system into either READY or 
STANDBY mode. The unit will only emit laser energy when the footswitch is pressed 
and the unit is set to READY mode. Values may be adjusted in both modes. In 
READY mode, values may be changed only when footswitch is released. 
 
READY Mode 
When entering READY mode, the system fan will turn ON and pressing the 
footswitch will activate laser radiation. There is 2 sec delay between switching to 
READY mode and the ability of the system to emit a laser beam.  This is evidenced 
by the delay in the appearance of the red aiming beam.