Device Error (Medium Priority) Alarms 
The GoSafe Lite Pendant alarms described below are automatically generated  
alarms designed to periodically let the Call Center know that the Pendant is working 
properly They are part of normal operation and are the result of no intensional action 
by the user. 
Alarm Event Type
Low Battery Capacity – 7150 MHB
Alarm Event 
Description
When the capacity / life of the battery in the Pendant has 
diminished materially and should be replaced, a Warning 
Message is displayed
Priority
Medium
Device Action
Charge cycle frequency increases, otherwise. Operates 
normally until battery is depleted.
Alarm Event 
Reported/Displayed 
Locally?
No.
Alarm Event 
Reported to Philips 
Lifeline?
Yes. A silent Maintenance Signal is sent to the Response 
Center by the device. Upon receiving the Maintenance 
Signal, the Response Center prioritizes the response and 
contacts the Subscriber to verify the status of  
their equipment.
Battery Requires Charging – 7150MHB
Alarm Event 
Description
When the rechargeable battery in the Pendant needs to be 
charged, a visual and audible warning Message is activated 
on the device.
Priority
Medium
Device Action
Operates normally until the battery is critically low.
Alarm Event 
Reported/ 
Displayed Locally?
Yes. The Indicator Light flashes orange and plays a single 
chime. This alarm will self cancel once the battery charge 
reaches an acceptable level.
Alarm Event 
Reported to Philips 
Lifeline?
No.