Lifeline Systems Inc. 7100MHB ユーザーズマニュアル
Questions? Call 1-800-635-6156
Depleted Battery – 7150 MHB
Alarm Event
Description
Description
When the charge level of the rechargeable battery is
critically low, a Warning Message is announced.
critically low, a Warning Message is announced.
Priority
Medium
Device Action
Device shuts down after the message is announced.
Alarm Event
Reported/
Displayed Locally?
Reported/
Displayed Locally?
Yes. Indicator Light flashes orange, Voice Prompt plays: “Your
help button is now out of power and will shut down. Please
charge your device as soon as possible”. This alarm will self
cancel once the battery charge reaches an acceptable level.
help button is now out of power and will shut down. Please
charge your device as soon as possible”. This alarm will self
cancel once the battery charge reaches an acceptable level.
Alarm Event
Reported to
Lifeline?
Reported to
Lifeline?
Yes. A silent Maintenance Signal is sent to the Response
Center by the End Device from anywhere wireless signal is
available on the AT&T wireless network. Upon receiving the
Maintenance Signal, the Response Center stores the status
information but no direct action is taken.
Center by the End Device from anywhere wireless signal is
available on the AT&T wireless network. Upon receiving the
Maintenance Signal, the Response Center stores the status
information but no direct action is taken.
Device Hardware Failure – 7150 MHB
Alarm Event
Description
Description
In the event that internal diagnostics indicate that the
Pendant it has a Hardware Error, a Warning Message is
displayed if possible based on the nature of the failure.
Pendant it has a Hardware Error, a Warning Message is
displayed if possible based on the nature of the failure.
Priority
Medium
Device Action
May have limited operation, depending on the failure
Alarm Event
Reported/Displayed
Locally?
Reported/Displayed
Locally?
No.
Alarm Event
Reported to Philips
Lifeline?
Reported to Philips
Lifeline?
Yes. A silent Maintenance Signal is sent to the Response
Center. Upon receiving the Maintenance Signal, the
Response Center prioritizes the response and contacts the
Subscriber to verify the status of their equipment.
Center. Upon receiving the Maintenance Signal, the
Response Center prioritizes the response and contacts the
Subscriber to verify the status of their equipment.