Transmitter (Digimatic) Operations Manual
2015.11
3.
Two-way communication (Wireless trigger)
Transmitter TXG series has reception function, also receiver
DR-200U/R has transmission function.
By this two-way
communication function, DR-200U/R can transmit trigger signal
to TXG so that TXG should transmit measured data to
DR-200U/R.
The merit is that without pushing DATA Sw. far
from DR-200U/R, wireless data transmission is available.
Trigger function is applied to maximum 16 transmitters.
16 measured data will be sent to DR-200U/R at once.
If you use DR-200mini as receiver, this function does not work.
3.1
Setting of wireless trigger
Let measuring equipment show “0.4”(number under 2
nd
decimal place is ignored.)
While pushing C Sw., push ID Sw. After completed setting,
buzzer sounds “Pi”.
*11 ID No. of transmitter which can
be triggered by DR-200U/R is limited from 0~15.
3.2
Wireless trigger start
Push DATA Sw., then buzzer sounds “Pi”.
Transmitter is ready to receive trigger signal from DR-200.
While waiting trigger signal, to save consumption of coin
battery, transmitter checks the signal 0.1ms every 0.5ms.
3.3
Trigger data transmission
Receiving trigger signal (data request signal) from
DR-200U/R, transmitter transmits measured data.
Maximum time from trigger till data receipt is about 4 sec.
for 1 transmitter. (For 16 transmitters 12 sec.)
It is because of intermittent trigger reception and automatic
retrial action between data trigger and transmission.
*12 Operation for receiver DR-200U/R is required.
Look at the operation manual of DR-200U/R.)
3.4
Wireless trigger halt and restart
Halt
Push C Sw. Then buzzer sounds “Pi”.
Transmitter stops waiting trigger signal.
If transmitter does not receive trigger signal 5 minutes, it
also automatically stops waiting trigger signal.
Restart Push DATA Sw.
3.5
Cancel of wireless trigger
Let measuring equipment show “0.5”(number under 2
nd
decimal place is ignored.)
While pushing C Sw., push ID Sw.
After completed setting, buzzer sounds “Pi”.
4.
Data reception acknowledgment from DR-200
By 2 way communication between transmitter & DR-200U/R,
buzzer of transmitter can be set to sound “PiPiPi” or “Pi”
which means data reception acknowledgment from DR-200.
Acknowledgments, buzzer sounds and their displays of
measuring equipment for setting/cancel are as follows.
Let measuring equipment show “0.8 / 0.6 / 0.9 / 0.7”
While pushing C Sw., push ID Sw.
Acknowledgment / Buzzer Sound
Display
Non-reception sound / “PiPiPi”
Set:0.8
Cancel:0.9
Normal reception sound / “Pi”
Set:0.6
Cancel:0.7
* 13 Shipping default: ”Cancel”
* 14 For longer battery life: Recommend you to “Cancel”
* 15 During wireless trigger, this function does not work.
* 16 Setting of normal reception sound is effective only if
non- reception sound is set.
* 17 Operation for receiver DR-200U/R is required.
(“Reply of reception acknowledgement”)
* 18
If you use DR-200mini as receiver, this function does not work.
Because of radio wave specifications TelemeasureⅡ is
incompatible with TelemeasureⅠ.
About radio wave specifications see operations manual
DR-200U/R.
Specifications shall be changed without notice.
Sadashin Kaname-cho Bldg. 1-4-11
Kaname-cho, Toshina-ku Tokyo 171-0043 JAPAN
TEL.+81-3-5995-3800 FAX.+81-3-5995-3831 www.vectrix.co.jp