11.2 Replacing the batteries of the HydrINS 2 / HydrINS 2 Mini flow meter
Apply the following procedure to replace the batteries in the head of the HydrINS 2 / HydrINS 2 Mini
flow meter:
flow meter:
1 – Open the head of the flow meter (see Para. 8.3).
2 – Disconnect the used battery and connect the new one.
3 – In Winfluid, click the HydrINS II or Mini HydrINS input channel. In the Sensor tab, click
Properties, and then Advanced. In the Maintenance tab, select the boxes for Batteries 1 and 2 in the
head, and then confirm by clicking OK. The battery level will then be reinitialised during probe
programming.
Properties, and then Advanced. In the Maintenance tab, select the boxes for Batteries 1 and 2 in the
head, and then confirm by clicking OK. The battery level will then be reinitialised during probe
programming.
4 – Click "Sensor" and then "program".
Figure 43: Reinitialisation of the battery levels in HydrINS 2 in Mode 1 in Winfluid
11.3 Replacing the display batteries
1 – Open the upper cover of the Display.
2 – Remove the used battery and insert the new battery.
3 – Close the cover of the Display.
4 – In Winfluid, click the HydrINS II or Mini HydrINS input channel. In the Sensor tab, click
Properties, and then Advanced. In the Maintenance tab, select the boxes for Battery 2, indicate the
capacity in Ah of the new battery, and then confirm by clicking OK. The battery level will then be
reinitialised during probe programming.
Properties, and then Advanced. In the Maintenance tab, select the boxes for Battery 2, indicate the
capacity in Ah of the new battery, and then confirm by clicking OK. The battery level will then be
reinitialised during probe programming.