 Operating Manual GMH 3851 
Page 14 of 22 
 
 
_____________________________________________________ 
_____________________________________________________________________________ 
 
8  Output 
The output can be used as serial interface (for GRS3100 or GRS3105 interface adapters) or as analogue 
output (0-1V). If none of both is needed, we suggest to switch the output off, battery life then is extended. 
8.1  Interface - Base Address ('Adr.') 
By using an electrically isolated interface converter USB3100, GRS3100 or GRS3105 (accessory) the device 
can be connected to a PC. 
With the GRS3105 it is possible to connect up to 5 instruments of the GMH3000 family to a single interface 
(please also refer to GRS3105-manual). As a precondition the base addresses of all devices must not be 
identical, make sure to configure the base addresses accordingly 
(refer menu point “Adr.” in chapter 4)..  
In order to avoid transmission errors, there are several security checks implemented (e.g. CRC). 
The following standard software packages are available for data transfer: 
  GSOFT3050:  
Operating and evaluation software for the integrated logger function 
  EBS20M/ -60M
20- / 60-channel software to record and display the measuring values 
  GMHKonfig
Software for a comfortable editing of the device (e.g. Material selection
…) 
In case you want to develop your own software we offer a GMH3000-development package including 
an universally applicable 32bit Windows functions library ('GMH3000.DLL') with documentation that can 
be used by all 'serious' programming languages. 
Programming examples for Visual Basic 6.0™, Delphi 1.0™, Testpoint™, Labview™ 
The Device has 2 Channels: 
Channel 1: Material-moisture in % and base-address 
Channel 2: Temperature 
Supported Interface-functions:  
Code 
Name/Function 
Code 
Name/Function 
x  X  0 
read nominal value 
x  x  202 
read unit of display 
x  X  3 
read system status 
x  x  204 
read decimal point of display 
x   
12 
read ID-no. 
x   
205 
read extended measuring type in display 
x  X  176 
read min measuring range 
x   
208 
read channel count 
x  X  177 
read max measuring range 
x  x  214 
read scale correction 
x  X  178 
read measuring range unit 
x  x  215 
set scale correction 
x  X  179 
read measuring range decimal point 
x  x  216 
read zero displacement 
x  X  180 
read measuring type  
x  x  217 
set zero displacement 
 
X  194 
set display unit 
x   
222 
read turn-off-delay 
x  X  199 
read measuring type in display 
x   
223 
Set turn-off-delay 
x  X  200 
read min. display range 
x   
240 
Reset 
x  X  201 
read max. display range 
x   
254 
read program identification 
 
 
The measuring and range values read via interface are always in the selected display 
unit (°C/°F)! 
8.2  Analogue Output 
– Scaling with DAC.0 and DAC.1 
With the DAC.0 and DAC.1 values the output can be rapidly scaled to Your efforts. 
Keep in mind not to connect low-resistive loads to the output, otherwise the output value will be wrong and 
battery life is decreased. Loads above ca 10kOhm are uncritical. 
If the display exceeds the value set by DAC.1, then the device will apply 1V to the output 
If the display falls below the value set by DAC.0, then the device will apply 0V to the output   
In case of an error (Err.1, Err.2, no sensor, etc.) the device will apply slightly above 1V to the output. 
Plug wiring: 
 
GND 
 
    
The 3
rd
 contact  has to be left floating! 
    Only stereo plugs are allowed!
 
+Uout 
Application in the glued timber construction acc. to DIN 1052-1 (MPA certified) 
The instrument with its curve h.460 (Fir) was certified by the MPA Stuttgart (Otto Graf institute) for 
applications in the glued timber construction according to DIN 1052-1 with the following equipment: 
measuring cable GMK38 
reciprocating piston electrode GHE91 (recommended) or impact electrode GSE91