Greisinger MATERIAL-MOISTURE DEVICE GMH 3851 605277 ユーザーズマニュアル

製品コード
605277
ページ / 22
H60.0.12.6C-13 
 Operating Manual GMH 3851 
Page 15 of 22 
 
 
_____________________________________________________ 
_____________________________________________________________________________ 
 
10 Fault and System Messages 
Display 
Meaning 
Remedy 
 
Blinking curve display: Displayed value is out of 
specified range (Wood: 8..40%u) 
Limited measuring precision!  
The display value is only usable as indication, 
not as measurement! 
 
low battery voltage, device will continue to work 
for a short time 
replace battery 
If mains operation: wrong voltage 
replace power supply, if fault continues to exist: 
device damaged 
 
low battery voltage 
replace battery 
If mains operation: wrong voltage 
Check/replace power supply, if fault continues 
to exist: device damaged 
No display 
or 
weird display 
 
Device does not 
react on keys 
low battery voltage 
replace battery 
If mains operation: wrong voltage 
Check/replace power supply, if fault continues 
to exist: device damaged 
system error 
Disconnect battery or power supply, wait some 
time, re-connect 
device defective 
return to manufacturer for repair 
---- 
Sensor error: no material connected (meas. 
Value below permissible range), no valid signal 
Connect meas. material 
charge at the probe, device will discharge (esp. 
at dry wood) 
Wait until probe has discharged 
Sensor broken or device defective  
return to manufacturer for repair 
Err.1 
Value exceeding measuring range 
Check: Is the value exceeding the measuring 
range specified? ->temperature too high! 
Wrong probe connected 
Check probe 
Probe or device defective 
return to manufacturer for repair 
Non-floating probe near the unshielded electrode  Insulate probe or measure at shielded electrode 
Err.2 
Value below display range 
Check: Is the value below the measuring 
range specified? -> temperature too low! 
Wrong probe connected 
Check probe  
Probe, cable or device defective 
return to manufacturer for repair 
Err.7 
system error 
return to manufacturer for repair 
11 Inspection of the accuracy / Adjustment Services  
Accuracy can be inspected with the testing adapter GPAD 38 (extra equipment). 
To check precision select 
material characteristic curve “.rEF”, choose display „%u“ and connect the testing adapter. The 
device must display the printed value for the GMH38xx. If the precision is bad, we suggest to send the device to the 
manufacturer for a new adjustment. 
12 Measuring precision 
 
Frozen wood cannot be measured! 
 
The measuring needles have to be fixed very well e.g. by means of a wrench 
Loose needles can disturb the measuring