T400 Calibration Procedures 
T400 Ozone Analyzer Operator’s Manual 
200 
 
To activate the dark calibration feature, press: 
SETUP X.X       
 PRIMARY SETUP MENU
 CFG DAS RNGE PASS CLK
MORE
EXIT
SAMPLE
RANGE=500.0 PPB          O3= XXXX
<TST
TST> CAL
SETUP
SETUP X.X         
SECONDARY SETUP MENU
 COMM VARS
DIAG
EXIT
SETUP X.X    
ENTER PASSWORD:818
 
8
 
 
1
 
 
8
ENTR
EXIT
DIAG                       SIGNAL I/O            
NEXT
 ENTR
EXIT
Press NEXT until ...
The DARK CAL procedure progresses automatically 
until ...
SETUP X.X       
DARK CAL 34% COMPLETE
SETUP X.X       
CALIBRATING DARK OFFSET
SETUP X.X       
DARK CALIBRATION
PREV NEXT
ENTR
EXIT
DARK CAL 
Successful?
Yes
No
SETUP X.X    
INVALID DARK CAL OFFS=XXXX.X MV
EXIT
 
06870C DCN6332