T400 Calibration Procedures
T400 Ozone Analyzer Operator’s Manual
200
To activate the dark calibration feature, press:
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SAMPLE
RANGE=500.0 PPB O3= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
DIAG SIGNAL I/O
NEXT
ENTR
EXIT
Press NEXT until ...
The DARK CAL procedure progresses automatically
until ...
SETUP X.X
DARK CAL 34% COMPLETE
SETUP X.X
CALIBRATING DARK OFFSET
SETUP X.X
DARK CALIBRATION
PREV NEXT
ENTR
EXIT
DARK CAL
Successful?
Yes
No
SETUP X.X
INVALID DARK CAL OFFS=XXXX.X MV
EXIT