T400 Ozone Analyzer Operator’s Manual 
T400 Calibration Procedures 
 
201 
9.5.2. O
3
 PHOTOMETER GAS FLOW CALIBRATION 
Note 
A separate flow meter is required for this procedure. 
To calibrate the flow of gas through the T400 analyzer’s optional photometer bench. 
1.  Turn OFF the T400 analyzer. 
2.  Attach the flow meter directly to the SAMPLE inlet port of the analyzer. 
3.  Turn the analyzer ON. 
4.  Perform the following steps: 
 