T400 Ozone Analyzer Operator’s Manual 
T400 Calibration Procedures 
 
203 
To calibrate the O
3
 Generator press: 
DIAG O3GEN        
    O3 GEN CAL [X]% COMPLETE
PREV NEXT
EXIT
SETUP X.X       
 PRIMARY SETUP MENU
 CFG DAS RNGE PASS CLK
MORE
EXIT
SAMPLE
RANGE=500.0 PPB          O3= XXXX
<TST
TST> CAL
SETUP
SETUP X.X         
SECONDARY SETUP MENU
 COMM VARS
DIAG
EXIT
SETUP X.X    
ENTER PASSWORD:818
 8
  1
  8
ENTR
EXIT
DIAG                       SIGNAL I/O            
NEXT
ENTR
EXIT
Continue pressing NEXT  until ...
DIAG O3GEN        
    CANCELLED
PREV NEXT
EXIT
Analyzer returns to the 
previous menu when the 
calibration operation is 
complete
DIAG              
       O3 GEN CALIBRATION
PREV NEXT
ENTR
EXIT
Display tracks 
the % of test 
completes
  