T400 Ozone Analyzer Operator’s Manual
T400 Calibration Procedures
203
To calibrate the O
3
Generator press:
DIAG O3GEN
O3 GEN CAL [X]% COMPLETE
PREV NEXT
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SAMPLE
RANGE=500.0 PPB O3= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
DIAG SIGNAL I/O
NEXT
ENTR
EXIT
Continue pressing NEXT until ...
DIAG O3GEN
CANCELLED
PREV NEXT
EXIT
Analyzer returns to the
previous menu when the
calibration operation is
complete
DIAG
O3 GEN CALIBRATION
PREV NEXT
ENTR
EXIT
Display tracks
the % of test
completes
06870C DCN6332