T400 Ozone Analyzer Operator’s Manual 
Troubleshooting & Service 
 
273 
Why does the ENTR button sometimes disappear on 
the Front Panel Display? 
During certain types of adjustments or configuration 
operations, the ENTR button will disappear if you 
select a setting that is nonsensical (such as trying to 
set the 24-hour clock to 25:00:00) or out of the 
allowable range for that parameter (such as selecting 
an DAS
 
Holdoff period of more than 20 minutes). Once 
you adjust the setting in question to an allowable 
value, the ENTR button will re-appear. 
How do I make the RS-232 Interface Work? 
See Section 6. 
How do I use the DAS? 
See Section 7. 
How do I make the instrument’s display and my data 
logger agree? 
This most commonly occurs when an independent 
metering device is used besides the data 
logger/recorded to determine gas concentration levels 
while calibrating the analyzer.  These disagreements 
result from the analyzer, the metering device and the 
data logger having slightly different ground levels.It is 
possible to enter a DC offset in the analog outputs to 
compensate.  This procedure is located in Section 
5.10.1.8 of this manual. Alternately, use the data 
logger itself as the metering device during calibration 
procedures. 
When should I change the Particulate Filter and how 
do I change it? 
The Particulate filter should be changed weekly.  See 
Section 11.3.1 for instructions on performing this 
replacement. 
When should I change the Sintered Filter and how do I 
change it? 
The Sintered Filter does not require regular 
replacement. Should its replacement be required as 
part of a troubleshooting or repair exercise, see 
Section 12.10.1 for instructions. 
When should I change the Critical Flow Orifice and 
how do I change it? 
The Critical Flow Orifice does not require regular 
replacement. Should its replacement be required as 
part of a troubleshooting or repair exercise, see 
Section 12.10.1 for instructions. 
How do I set up and use the Contact Closures (Control 
Inputs) on the Rear Panel of the analyzer? 
See Section 3.3.1.6. 
Can I automatically calibrate or check the calibration of 
my analyzer? 
Any analyzer into which a Zero/Span Valve Option can 
be automatically calibrated using the instrument’s 
AutoCal Feature. Be aware that while the AutoCal 
feature can be used with the IZS Option to perform 
Calibration Checks, The IZS should never be used to 
perform Calibrations. See Section 9.4 for instructions 
on setting up and activating the AutoCal feature. 
How often should I rebuild the Sample Pump on my 
analyzer? 
The diaphragm of the Sample Pump should be 
replaced annually. A sample rebuild kit is available.  
See Appendix B of this manual for the part number of 
the pump rebuild kit.  Instructions and diagrams are 
included with the kit. 