9600 Compressor Installation, Operation, and Maintenance
P/N 8040540
3-3
C
HELIX  TECHNOLOGY  CORPORATION
-
TI
CRYOGENICS
Supply and Return Water Line Connections
NOTE: The water used for cooling the Compressor must meet the 
specifications outlined in Section 1 - Compressor Description
.
Hard Water Lines
1. Apply a light coating of standard plumbing thread sealant to the 
hard line pipe threads.
2. Install the Supply hard line into the INPUT connection on the rear 
panel of the Compressor.  Tighten the fitting by hand.
3. Install the Return hard line into the OUTPUT connection on the 
rear panel of the Compressor.  Tighten the fitting by hand.
CAUTION
Do not overtighten the ferrules.  Damage to the input and output      
connector threads may occur.
4. Using a wrench, tighten the fittings.
5. Allow water to flow and check for leaks at the rear of the 
Compressor.
Flexible Water Lines
1. Apply a light coating of standard plumbing thread sealant to the 
barbed fitting threads.
2. Install the barbed fittings into the INPUT and OUTPUT 
connections on the rear panel of the Compressor.
CAUTION
Do not overtighten the barbed fittings.  Damage to the INPUT and 
OUTPUT connector threads may occur.
3. Using a wrench, tighten the barbed fittings.
4. Connect the Supply flexible water line to the INPUT barbed fitting 
and secure with a hose clamp.
5. Connect the Return flexible water line to the OUTPUT barbed       
fitting and secure with a hose clamp.
6. Allow water to flow and check for leaks at the rear of the           
Compressor.