EN-20
CD-11 Indicator
Press the G/N/T/
Menu 
button, LStOFF is displayed.  This permits locking the
SEtuP menu. LStOFF is unlocked, LSt On is read only (locked).  This menu is
hidden if the CAL jumper is off.
Press the     PRINT/
Units 
button for selections LSt ON or LStOFF.
Press the G/N/T/
Menu 
button, LrdOFF displayed.  This permits locking the rEAd
menu. LrdOFF is unlocked, Lrd On is read only (locked).
Press the PRINT/
Units 
button for selections Lrd On or LrdOFF.
Press the G/N/T/
Menu
     button, LPtOFF is displayed.  This permits locking the
Print menu. LPtOFF is unlocked, LPtOn is read only (locked).
Press the PRINT/
Units 
button for selections LPtOn or LPtOFF.
Press the G/N/T/
Menu 
button, LCLOFF is displayed.  This permits locking the
Calibration menu. LCLOFF is unlocked, LCL On is read only (locked).  This
menu is hidden if the CAL jumper is off.
Press the PRINT/
Units 
button for selections LCL On or LCLOFF.
Press the G/N/T/
Menu 
     button to end this block, SAVE is displayed.
Press the G/N/T/
Menu 
button to save the lockout menu settings.  The next
menu Quit is displayed.
or
Press the PRINT/
Units
 button to go back to the lockout menu without saving.
After saving the lockout menu settings, press the PRINT/
Units 
button to go to
CAL or press the G/N/T/
Menu 
button to return to a weighing mode.
NOTE:  At this point, the Indicator must be calibrated and the jumper removed
from the CAL connector in order to lock the menus. The top cover of the Indica-
tor should be free to gain access to the CAL jumper.
2.6.7 Lockout Menu (Cont.)
Stable
Gross
     Brutto
> 0 <
> 0 <
> 0 <
> 0 <
> 0 <
*