Ohaus ohaus indicator cd-11 ユーザーズマニュアル

ページ / 120
EN-22
CD-11 Indicator
Procedure
LINEARITY CALIBRATION
With the Indicator ON, press and hold the     button G/N/T/
Menu 
until MEnU is
displayed.  When the      G/N/T/
Menu
 button is released, CAL is displayed.
Press the G/N/T/
Menu
      button, SPAN Y is displayed.
Press the PRINT/
Units
 button, Lin  Y is displayed.
Press the G/N/T/
Menu
     button, -C- is displayed.  The scale base 
MUST
 be stable
during this period as it establishes a zero point.  After a few seconds, the dis-
play flashes LIN CP twice and the requested weight value is displayed. The
sample illustration indicates a 2kg center point for a 5kg scale.
Place the indicated mass on the platform.  Keep the platform stable during this
period.
Press the G/N/T/
Menu
     button, -C- is displayed.  The scale base 
MUST
 be stable
during this period as it establishes a center point.  After a few seconds, the dis-
play flashes FULLCP twice and the requested weight value is displayed.
Place the indicated mass on the platform and press the G/N/T/
Menu
     button -C-
is displayed.
If linearity calibration was successful, the calibration weight is displayed and
the calibration data is saved automatically.  If unsuccessful, refer to the trouble-
shooting section.
Remove calibration masses from platform.
Gross
      Brutto
NOTE:  If the Indicator is to be used for legal for trade or legally controlled applications, it must be calibrated
and the jumper removed from the CAL connector in order to lock out the menus. The top cover of the Indicator
should be free to gain access to the CAL jumper.  Refer to section 3.1 for sealing for legal for trade use.
Stable