EN-23
CD-11 Indicator
GEOGRAPHICAL FACTOR (For Europe Only)
Press the G/N/T/
Menu
 button, gEO 19 is displayed.  This is the current geo-
graphical adjustment value.
The geo factor includes settings from 0 to 31 and is used to compensate for
varying gravity at different geographical areas (complete geographical adjust-
ment settings are listed in the following table).
NOTE:
Only an authorized manufacturer’s representative or certified verification
personnel may make these changes.
Changing the geographical setting alters the calibration values.
Press the Print/
Units
 button until the desired geographical adjustment value is
reached.  The factory default setting is gEO 19
SAVE
Press the G/N/T/
Menu
 button to end this block, SAVE is displayed.
Press the G/N/T/Menu button to save the geographical factor setting.  The next
menu SEt uP is displayed.
or
Press the Print/
Units
 button to go back to the CAL menu without saving
changes to the geographical setting.
QUIT
To exit from the CAL menu, press the Print/
Units
 button to advance to Quit.
Then press the G/N/T/
Menu
 button to go back to the weighing mode.