OPERATING THE PATIENT LIFT
24
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: Use the primary emergency release 
first. The secondary emergency release is only a 
back‐up to the primary emergency release.
In cases where the primary release is either 
not functioning or unreachable, a 
secondary emergency release may be used.
To activate the secondary release, pull up 
on the EMERGENCY ring and push down 
on the boom at the same time.
FIGURE 3.3   Secondary Emergency Release
Performing an Emergency Stop
NOTE: For this procedure, refer to FIGURE 3.4.
Press the RED/ORANGE emergency button on the control box to stop the boom assembly 
and patient from raising or lowering.
Rotate the RED/ORANGE emergency stop button clockwise to disengage the emergency 
stop.
FIGURE 3.4   Performing an Emergency Stop
Charging the Battery
NOTE: For this procedure, refer to FIGURE 3.5 on page 25.
NOTE: Invacare recommends the battery be recharged daily to prolong battery life. 
NOTE: An audible alarm will sound (horn will beep) when battery is low.
1. Lift up on the handle on the back of the battery.
2. Lift the battery up and out of the control box.
Pull UP on 
EMERGENCY 
Ring
Push DOWN 
on Boom
Control Box
Press IN to 
Stop Boom
Rotate CLOCKWISE to 
Disengage Emergency Stop