Invacare R121 ユーザーズマニュアル
OPERATING THE PATIENT LIFT
23
Opening/Closing the Legs
To open the legs, press the legs open
button.
button.
To close the legs, press the legs closed
button.
button.
FIGURE 3.1 Pendant Buttons
Activating a Mechanical Emergency Release
Primary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.2.
All patient lift actuators are equipped with
a mechanical emergency release. The
mechanical release will enable the actuator
to retract without power. The actuator will
only retract if a load is applied and the
mechanical emergency release button is
pulled. The red/orange emergency release
button has emergency written upon it.
a mechanical emergency release. The
mechanical release will enable the actuator
to retract without power. The actuator will
only retract if a load is applied and the
mechanical emergency release button is
pulled. The red/orange emergency release
button has emergency written upon it.
To activate the primary emergency release,
insert a pen into the hole labeled
emergency on the control box of the
patient lift and push down on the boom at
the same time.
insert a pen into the hole labeled
emergency on the control box of the
patient lift and push down on the boom at
the same time.
FIGURE 3.2 Primary Emergency Release
Legs Closed
Button
Boom Up
Button
Legs Open
Button
Boom Down
Button
EME
RGENCY
EMERGENCY
Emergency
Release
Button
Control
Box