Invacare R121 ユーザーズマニュアル

ページ / 52
OPERATING THE PATIENT LIFT
22
SECTION 3—OPERATING THE 
PATIENT LIFT
Introduction
 WARNING
DO NOT attempt any transfer of a patient without approval of the patient's 
physician, nurse, or medical assistant. Thoroughly read the instructions in this 
owner's manual, observe a trained team of experts performing the lifting 
procedures and then perform the entire lift procedure several times with proper 
supervision and a capable individual acting as a patient.
The legs of the patient lift must be in the maximum open position for optimum sta-
bility and safety. If it is necessary to close the legs to maneuver the patient lift under 
a bed, close the legs only as long as it takes to position the patient lift over the 
patient and lift the patient off the surface of the bed. When the legs of the patient 
lift are no longer under the bed, return the legs to the maximum open position.
NOTE: Invacare recommends that two assistants be used for all lifting preparation and 
transferring to/from procedures; however, the patient lift can be operated with one assistant. The 
use of one assistant is based on the evaluation of the health care professional for each individual 
case.
Using the Pendant Buttons
 WARNING
DO NOT lock the rear casters of the patient lift when lifting an individual. Locking 
the rear casters could cause the patient lift to tip and endanger the patient and 
assistants.
NOTE: For this procedure, refer to FIGURE 3.1 on page 23.
The pendant is used to raise/lower the boom or to open/close the legs of the base for 
stability when lifting a patient
Raising/Lowering the Boom
To raise the patient lift, press the boom up button (up arrow) to raise the boom and the 
patient.
To lower the patient lift, press the boom down button (down arrow) to lower the boom 
and the patient.