370 VT1422A Command Reference  
[SENSe]
Related Commands: STRAIN:UNST
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
Usage
STRAIN:UNST?  (@107)
query unstrained voltage for channel 7
enter statement here
returns the unstrained voltage set by 
STR:UNST
[SENSe:]TOTalize:RESet:MODE
[SENSe:]TOTalize:RESet:MODE <select>,<ch_list
sets the mode for resetting 
totalizer channels in <ch_list>.
Parameters
Comments
In the INIT mode the total is reset only when the INITiate command is executed. 
In the TRIGger mode the total is reset every time a new scan is triggered.
If the channels specified are not on a Frequency/Totalize SCP, an error will 
be generated.
Related Commands: SENS:FUNC:TOT, INPUT:POLARITY
*RST Condition: SENS:TOT:RESET:MODE INIT
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENS:TOT:RESET:MODE TRIG,(@134)
totalizer at channel 34 resets at each 
trigger event
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
select
discrete (string)
INIT | TRIGger
seconds
ch_list
string
100 - 163
none