II Software for Micro Motion
®
Transmitters
Installation and Setup
To make the software connection from ProLink II to your transmitter.
1. Ensure that your PC is connected to the transmitter, according to one of the methods described
2. Start ProLink II software.
3. From the
Connection
menu, click on
Connect to Device
.
4. Specify connection parameters:
•
Use the
Protocol
parameter to specify your connection type. For HART/Bell 202
connections using the VIATOR USB HART Interface, enable
Converter Toggles RTS
.
•
Set
Serial Port
to the PC COM port you are using to connect to the transmitter.
•
If you are making a service port or HART/Bell 202 connection, default values are used for
all remaining connection parameters.
all remaining connection parameters.
•
If you are connecting to MVD Direct Connect, set the remaining connection parameters to
any of the supported settings listed in Table 2-13.
any of the supported settings listed in Table 2-13.
•
For all other connection types:
-
For Model 2400S transmitters, set the address to the Modbus address configured for
your transmitter.
your transmitter.
Note: Due to the transmitter’s auto-detection feature, other connection parameters are not required.
-
For all other transmitters, set the remaining connection parameters to the values
configured in your transmitter.
configured in your transmitter.
Table 2-16
Default communication parameters for RFT97xx and IFT97xx transmitters
Default values
Transmitter
Physical layer
Protocol
Baud
Data bits
(1)
(1) ProLink II automatically sets data bits appropriately for the configured protocol. Even though a data bits parameter may be
configured in the transmitter, you do not need to configure it in ProLink II.
Stop bits
Parity
Address
IFT9701/9703
(2)
(2) IFT9701/9703 communication parameters are not configurable. The settings shown here are always in effect.
Bell 202
(3)
(3) Connection to primary mA output.
HART
1200
8
1
odd
0
RFT9712
Bell 202
HART
1200
8
1
odd
0
RS-485
(4)
(4) Connection to RS-485 terminals.
HART
1200
8
1
odd
0
RFT9739 v2
Bell 202
HART
1200
8
1
odd
0
RS-485
HART
1200
8
1
odd
0
RFT9739 v3
Bell 202
HART
1200
8
1
odd
0
RS-485
(5) Dip switch settings on the transmitter are used to select either Std. comm or User defined.
• Std. comm
Modbus RTU
9600
8
1
odd
1
• User defined
HART
1200
8
1
odd
0