To make the software connection from ProLink II to your transmitter.
1. Ensure that your PC is connected to the transmitter, according to one of the methods described 
2. Start ProLink II software.
3. From the 
Connection
 menu, click on 
Connect to Device
.
4. Specify connection parameters:
Use  the 
Protocol
 parameter to specify your connection type. For HART/Bell 202 
connections using the VIATOR USB HART Interface, enable 
Converter Toggles RTS
.
Set 
Serial Port
 to the PC COM port you are using to connect to the transmitter.
If you are making a service port or HART/Bell 202 connection, default values are used for 
all remaining connection parameters.
If you are connecting to MVD Direct Connect, set the remaining connection parameters to 
any of the supported settings listed in Table 2-13.
For all other connection types:
-
For Model 2400S transmitters, set the address to the Modbus address configured for 
your transmitter. 
Note: Due to the transmitter’s auto-detection feature, other connection parameters are not required.
-
For all other transmitters, set the remaining connection parameters to the values 
configured in your transmitter.
Table 2-16
Default communication parameters for RFT97xx and IFT97xx transmitters
Default values
Transmitter
Physical layer
Protocol
Baud
Data bits
(1)
(1) ProLink II automatically sets data bits appropriately for the configured protocol. Even though a data bits parameter may be 
configured in the transmitter, you do not need to configure it in ProLink II.
Stop bits
Parity
Address
IFT9701/9703
(2)
(2) IFT9701/9703 communication parameters are not configurable. The settings shown here are always in effect.
Bell 202
(3)
(3) Connection to primary mA output.
HART
1200
8
1
odd
0
RFT9712
Bell 202
HART
1200
8
1
odd
0
RS-485
(4)
(4) Connection to RS-485 terminals.
HART
1200
8
1
odd
0
RFT9739 v2
Bell 202
HART
1200
8
1
odd
0
RS-485
HART
1200
8
1
odd
0
RFT9739 v3
Bell 202
HART
1200
8
1
odd
0
RS-485
(5) Dip switch settings on the transmitter are used to select either Std. comm or User defined.
• Std. comm
Modbus RTU
9600
8
1
odd
1
• User defined
HART
1200
8
1
odd
0