18
Check the 3M
TM 
Ranger
TM
 system website to ensure you have the most recent version of this document. 
www.rangerfluidwarming.com reorder #202524A
Ranger Irrigation Fluid Warming, Model 247 Maintenance Log
The operating temperature and alerts should be tested at least every 6 months or according to 
institutional protocol. 
 
m
AintenAnCe
 A
Ction
 p
eRfoRmed
 
model
  ___________  
Sn
  _______
Date  ________  
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(48°C [±2] on Ranger unit display) 
______°C
 
Secondary 
(50°C [±2])
 ______°C
Date  ________  
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(48°C [±2] on Ranger unit display) 
______°C
 
Secondary 
(50°C [±2])
 ______°C
Date  ________  
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(48°C [±2] on Ranger unit display) 
______°C
 
Secondary 
(50°C [±2])
 ______°C
Date  ________  
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(48°C [±2] on Ranger unit display) 
______°C
 
Secondary 
(50°C [±2])
 ______°C
Date  ________  
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(48°C [±2] on Ranger unit display) 
______°C
 
Secondary 
(50°C [±2])
 ______°C
Date  ________  
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(48°C [±2] on Ranger unit display) 
______°C
 
Secondary 
(50°C [±2])
 ______°C
M
od
el
 2
47