AngioDynamics UK Limited AS0670 사용자 설명서
DELTA 15/30
DELTA 15/30 Operator Manual
Page 2.3
Version B.2
WARNINGS
US Federal Law restricts the use of this device to sale by or on the order of a
physician.
Intended for use only by trained physicians/surgeons familiar with laser procedures.
physician.
Intended for use only by trained physicians/surgeons familiar with laser procedures.
Use of controls or adjustments or performance of procedures other than those
specified herein may result in hazardous radiation exposure.
specified herein may result in hazardous radiation exposure.
This product must be stored at temperatures between 0
°
C and +55
°
C. If stored at
temperatures below 10°C for a period of time, the l aser requires up to 12 hours to
acclimatise, prior to operation.
Failure to observe this could result in invalidation of the guarantee.
acclimatise, prior to operation.
Failure to observe this could result in invalidation of the guarantee.
The laser is not designed to operate at temperatures below 10
°
C.
The Optical Power Calibration meter, if fitted, is to be used only to calibrate a fiber or
spot handpiece in accordance with the instructions detailed in this manual.
Under no circumstances may it be used as a beam dump, as this will result in
damage to the instrument.
spot handpiece in accordance with the instructions detailed in this manual.
Under no circumstances may it be used as a beam dump, as this will result in
damage to the instrument.
This product contains a lithium battery, which should only be replaced by authorised
service personnel.
Replace the battery only with the same or equivalent type. Dispose of used batteries
according to the manufacturer’s instructions and local disposal requirements.
service personnel.
Replace the battery only with the same or equivalent type. Dispose of used batteries
according to the manufacturer’s instructions and local disposal requirements.