EMC Warning
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided.
The DIOMED DELTA 15/30
put into service according to the EMC information provided.
The DIOMED DELTA 15/30
may be interfered with by other equipment, even if that other equipment
complies with CISPR emission requirements.
Portable and Mobile RF communications equipment can affect medical electrical equipment.
The DIOMED DELTA 15/30 should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the DIOMED DELTA 15/30 should be observed to verify normal
operation in the configuration in which it will be used.
If Electromagnetic interference is experienced, relocate or re-orientate the DIOMED DELTA 15/30
Portable and Mobile RF communications equipment can affect medical electrical equipment.
The DIOMED DELTA 15/30 should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the DIOMED DELTA 15/30 should be observed to verify normal
operation in the configuration in which it will be used.
If Electromagnetic interference is experienced, relocate or re-orientate the DIOMED DELTA 15/30
or
the other equipment.
Accessories, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the DIOMED DELTA 15/30
transducers and cables sold by the manufacturer of the DIOMED DELTA 15/30
as replacement
parts for internal components, may result in increased emissions or decreased immunity of the
DIOMED DELTA 15/30. ‘Immunity’ is the ability of a device to function normally when operated
in the presence of electromagnetic radiation.
DIOMED DELTA 15/30. ‘Immunity’ is the ability of a device to function normally when operated
in the presence of electromagnetic radiation.
The following Cables are compatible with the DIOMED DELTA 15/30
:
•
Foot Switch Assembly
Diomed part no. AS1/A0/0002
•
Remote Interlock Lead (if required).
Lemo connector (supplied with laser): Diomed part no. CON/51/0003
Ferrite sleeve (Farnell part no. 898-454).
Screened twisted pair cable 7/0.2 (Farnell part no. 140-457).
Ferrite sleeve (Farnell part no. 898-454).
Screened twisted pair cable 7/0.2 (Farnell part no. 140-457).
Note: The ferrite sleeve should be fitted to the remote interlock cable, at a maximum of 75 mm
from the Lemo connector.
from the Lemo connector.
•
Remote Interlock Bypass (supplied with laser).
Diomed part no. AS1/A3/0024
Diomed part no. AS1/A3/0024
•
IEC Mains Lead (supplied with Laser)
Diomed part numbers: CBL/02/0002, CBL/02/0040, CBL/02/0042, CBL/02/0046,
Diomed part numbers: CBL/02/0002, CBL/02/0040, CBL/02/0042, CBL/02/0046,
CBL/02/0051, CBL/02/0063.